oWorldwide competence Home | News | Products | About us | Support | Contact us
USA
maMoS-100 / maMoS-200

Pictures

Stationary Gas Monitor
f
or 1 or 2 gas components

Low-cost gas monitor, either with electrochemical or IR sensor, for permanent display of one or two gas components. Comes complete with filter, peltier dryer and condensate pump. maMoS is equipped with analog outputs and a PC interface.

Description
maMoS is a high quality gas monitor. The device is produced using the latest technology. Due to the modular design of maMoS it is possible to meet the individual client's expectations by adjusting it to performing particular measurements, such as concentration of various gases, pressure, temperature and a number of boiler parameter measurements.
The monitor is in particular designed for flue (boiler) gas analysis, but it can be used for different purposes (such as producing biogas from garbage and animal excrements) as well.
Each maMoS monitor, according to its configuration, enables the user to measure concentrations of one or more gases. Additionally the device can present the measurement results on the liquid crystal displays and up to eight analogue outputs. Independently of the built-in measurement equipment there is also the possibility of reading the measurement results and changing the settings with the help of the computer programme provided.

Product info
Technical data
Measuring ranges
Folder maMoS-100 (80 kB)
Folder maMoS-200 (80 kB)
Manual (5 MB)
Control software (15 MB)
Measurement of gas concentrations
  • Gas concentrations measured with NDIR sensors: CO, CO2, CH4, N2O
  • Gas concentrations measured with partial pressure sensor: O2
  • Gases measured electrochemically: CO, NO, NO2, SO2, H2S, H2
Other measurements
  • Atmospheric pressure
  • Differential pressure
  • Temperature using thermocouples
  • Temperature using thermistors Pt-500
Application possibilities Software capabilities
  • Biogas plants
  • Landfills
  • Greenhouses
  • Stationary monitoring of burner systems
  • Process control in various industries (cement, glass, incineration etc.)
  • Monitoring for air quality control, for example: CO2 (0...2000ppm) or CH4 (0...5%)
  • Stationary operation: Zeroing and measurement times user programmable
  • Permanent automatic check of the instruemnt
  • Compensation of cross sensitivity and temperature drift of gas sensors
Processing and presentation of measured data Hardware capabilities
  • Gas concentrations shown as ppm or mg/m3
  • All results shown as instantaneous values on display
  • Averaging of all measured values. Averaging time: 2 - 60 seconds for IR sensors and 2 - 30 seconds for EC sensors
  • Optional: Datalogger with 256MB MMC (Memory card)
  • Powerful PC program for analyser settings and data communication
  • Optional calculation of combustion parameters: Lambda, qA, Eta
  • Optional calculation of NOx from NO or NO and NO2
  • ABS housing for wall-mounting with internal ventilation
  • Microprocessor controlled
  • Correction of IR sensors with zero and one calibration gas possible
  • IR sensors (CO2 and CH4) with short reaction time (T90 < 15 seconds)
  • Integrated Peltier sample conditioner with cndensate pump and filter
  • Solenoid valve for automatic ventilation during zeroing
  • Membrane pump for sample extraction
  • 4 current (0/4 - 20 mA) and 4 voltage (0 - 10 V). Freely allocated to the measurment channels
  • Optional integrated clock/calendar
  • Up to 3 LCD displays. Freely allocated to the measurment channels
  • RS232C or RS485 connection for data transfer
  • Optional: Converter for connection to USB or Ethernet
  • PC programme for quick and simple configuration of hte instrument and data transfer
A