| Component |
Method |
Range |
Resolution |
Accuracy |
Detection Level |
Time (T90) |
| Standard gas measurements |
| O2 - Oxygen, Volume concentration |
partial pressure sensor |
0..25 % |
0.01% |
±0.2 %
or 2 % rel. |
0.01% |
45 s |
| CO2 Carbon dioxide, Volume concentration |
Calculated from O2 volume concentration |
0..25 % |
0.01 % |
±0.2 %
or 2% rel. |
0.01% |
45 s |
|
| CO Carbon monoxide, Volume concentration |
electrochemical sensor |
0..2000 ppm |
0.1 or 1 ppm
as set |
±5 ppm
or 5 % rel. |
0.1 ppm |
45 s |
| CO - Carbon monoxide, Volume concentration |
electrochemical sensor |
0..20,000 ppm |
1 ppm |
±5 ppm
or 5% rel. |
1 ppm |
45 s |
| CO Carbon monoxide, Volume concentration |
electrochemical sensor |
0..10 % |
10 ppm
or 0.001 % |
±50 ppm
or 5 % rel. |
10 ppm |
45 s |
| COmg Carbon monoxide, mass concentration |
Calculated from CO volume concentration |
0... |
1 mg/Nm3 |
± 5 mg/Nm3
or 5 % rel. |
1 mg/Nm3 |
45 s |
| COrel Carbon monoxide, concentration relative to O2 |
Calculated from CO and O2 volume concentrations |
0... |
1 mg/Nm3 |
± 5 mg/Nm3
or 5 % rel. |
1 mg/Nm3 |
45 s |
|
| NO Nitric oxide, Volume concentration |
electrochemical sensor |
0..1000 ppm |
0.1 or 1ppm
as set |
±5 ppm
or 5 % rel. |
1 ppm |
45 s |
| NOmg Nitric oxide, mass concentration |
Calculated from NO volume concentration |
0... |
1 mg/Nm3 |
±10 mg/Nm3
or 5 % rel. |
2 mg/Nm3 |
45 s |
| NOrel Nitric oxide, concentration relative to O2 |
Calculated from NO and O2 volume concentrations |
0... |
1 mg/Nm3 |
±10 mg/Nm3
or 5 % rel. |
2 mg/Nm3 |
45 s |
| Optional electrochemical sensors |
| NO2 Nitrogen dioxide, Volume concentration |
electrochemical sensor |
0...200 ppm |
0.1 or 1 ppm
as set |
± 5 ppm
or 5 % rel. |
1 ppm |
45 s |
| NO2mg Nitrogen dioxide, mass concentration |
Calculated from NO2 volume concentration |
0... |
1 mg/Nm3 |
± 10 mg/Nm3
or 5 % rel. |
2 mg/Nm3 |
45 s |
| NO2rel - Nitrogen dioxide concentration, relative to O2 |
Calculated from NO2 and O2 volume concentrations |
0... |
1 mg/Nm3 |
±10 mg/Nm3
or 5 % rel. |
2 mg/Nm3 |
45 s |
|
| SO2 Sulphur dioxide, Volume concentration |
electrochemical sensor |
0...2000 ppm |
0.1 or 1 ppm
as set |
±5 ppm
or 5 % rel. |
1 ppm |
45 s |
| SO2mg Sulphur dioxide, mass concentration |
Calculated from SO2 volume concentration |
0... |
1 mg/Nm3 |
± 15 mg/Nm3
or 5 % rel. |
3 mg/Nm3 |
45 s |
| SO2rel Sulphur dioxide concentration, relative to O2 |
Calculated from O2 and SO2 volume concentrations |
0... |
1 mg/Nm3 |
±15 mg/Nm3
or 5 % rel. |
3 mg/Nm3 |
45 s |
|
| H2S Hydrogen sulphide, Volume concentration |
electrochemical sensor |
0...1000 ppm |
0.1 or 1 ppm
as set |
±5 ppm
or 5 % rel. |
1 ppm |
45 s |
| H2Smg Hydrogen sulphide, mass concentration |
Calculated from H2S volume concentration |
0... |
1 mg/Nm3 |
±15 mg/Nm3
or 5 % rel. |
3 mg/Nm3 |
45 s |
| H2Srel Hydrogen sulphide, mass concentration relative to O2 |
Calculated from H2S and O2 volume concentrations |
0... |
1 mg/Nm3 |
±15 mg/Nm3
or 5 % rel. |
3 mg/Nm3 |
45 s |
|
| H2 Hydrogen, volume concentration |
electrochemical sensor |
0...2000 ppm |
0.1 or 1 ppm
as set |
±5 ppm
or 5 % rel. |
1 ppm |
45 s |
| H2mg Hydrogen, mass concentration |
Calculated from H2 volume concentration |
0... |
1 mg/Nm3 |
±15 mg/Nm3
or 5 % rel. |
3 mg/Nm3 |
45 s |
| H2rel Hydrogen, mass concentration relative to O2 |
Calculated from H2 and O2 volume concentrations |
0... |
1 mg/Nm3 |
±15 mg/Nm3
or 5 % rel. |
3 mg/Nm3 |
45 s |
| Optional IR sensors |
| CO Carbon monoxide, Volume concentration |
IR Sensor |
0..5 % 0...100% |
0.01%, 0.1% |
±0.3 % abs.,
or 3 % rel. |
0.01%, 0.1% |
45 s |
| CO2 Carbon dioxide, Volume concentration |
IR Sensor |
0..5 % 0...100% |
0.01%, 0.1% |
±0.3 % abs.,
or 3 % rel. |
0.01%, 0.1% |
45 s |
| CH4 Methane. Volume concentration |
IR Sensor |
0...5%, 0...100% |
0.01%, 0.1% |
±0.3 % abs.,
or 3 % rel. |
0.01%, 0.1% |
45 s |
| Measured temperatures |
| Tgas Temperature of the flue gases |
Thermocouple |
-20..2000°C |
1°C |
±2 °C
or 1. 5 % rel. |
1 °C |
30 s |
| Tamb Ambient temperature |
Thermistor |
-20..100°C |
0.1°C |
±1 °C |
1 °C |
30 s |
| Further values |
| Atmospheric pressure |
DMS bridge |
800 hPa ... 1200 hPa |
1 hPa |
±10 hPa |
1 hPa |
10 s |
| Pressure/draught |
DMS bridge |
-25 hPa ... +25 hPa |
0.1 or 1 Pa as set |
±2 Pa
or 5 % rel. |
0.1 Pa |
10 s |
| Diff. pressure |
DMS bridge |
-20 hPa ... +20 hPa |
0.1 or 1 Pa as set |
±2 Pa
or 5 % rel. |
0.1 Pa |
10 s |
| V Flow velocity |
Pitot tube |
1..30 m/s |
0.1 m/s |
0.5 m/s
or 2 % rel. |
1 m/s |
10 s |
| Calculated combustion parameters |
| TI (CO/CO2-Toxic Index) |
Calculated |
0...1 % |
0.01 % |
0.01 % |
|
|
| Lambda Excess air |
Calculated |
1..10 |
0.01 |
0.01 |
|
|
| qA Combustion loss |
Calculated |
0..100 % |
0.1 % |
0.1 % |
|
|
| Eta -Efficiency |
Calculated |
0..100 % |
0.1 % |
0.1 % |
|
|
| Parameter |
Description |
| Size |
W x L x H: 600 x 635 x 516 mm |
| Weight without probe |
48 kg |
|
| Display size |
Graphic LCD with backlighting. Contrast set by user, 320 x 240 pixels |
| Data logger |
MMC 256 MB |
| Interface |
RS232C |
|
| Supply |
Mains 230 VAC 50 ÷ 60Hz |
| Gas pump |
Membrane pump, set to 90 l/h |
| condensate pump |
peristaltic pump, electronically controlled |
|
| Sampling |
|
| Probe length |
300mm (other lengths available) |
| Sample hose |
Heated hose, 3 m |
| Gas filter |
Heated filter 20 µm |
|
| Analogue outputs |
|
| Voltage |
8 channels: 0...10 V |
| Current |
8 channels: 0/4...20 mA |
|
| Operating temperature |
10°C ÷ 40°C |
| Storage temperature |
-20 °C ÷ +55 °C |
| Humidity |
5 - 90 %, non-condensing |
| Humidity |
5 - 90 %, non-condensing |
| Parameter |
Description |
| Measuring method |
NDIR light absorption, 1 channel with light modulation |
| Averaging time |
2 ÷ 60 s freely programmable |
| Time T90 |
< 45 s |
| Measuring range |
Begin and end can be freely programmed within the nominal range |
| Accuracy |
±5 ppm abs.,or 5 % rel. |
| Calibration |
10 point calibration, stored in EEPROM |
| Recalibration |
not necessary |
|
Parameter
|
Description only for electrochemical sensors |
| Measuring method |
electrochemical |
| Averaging time |
2÷30 s freely programmable |
| Time T90 |
< 45 s |
| Measuring range |
Begin and end can be freely programmed within the nominal range |
| Accuracy |
± 5 ppm or 5 % rel. |
| Calibration |
2 point calibration, variable |
| Recalibration |
At least every 12 months |