cms-7
Stationary Gas Analyzer
Technical data
Component Method Range Resolution Accuracy Detection Level Time (T90)
Standard gas measurements
O2 - Oxygen, Volume concentration partial pressure sensor 0..25 % 0.01% ±0.2 %
or 2 % rel.
0.01% 45 s
CO2 – Carbon dioxide, Volume concentration Calculated from O2 volume concentration 0..25 % 0.01 % ±0.2 %
or 2% rel.
0.01% 45 s
CO – Carbon monoxide, Volume concentration electrochemical sensor 0..2000 ppm 0.1 or 1 ppm
as set
±5 ppm
or 5 % rel.
0.1 ppm 45 s
CO - Carbon monoxide, Volume concentration electrochemical sensor 0..20,000 ppm 1 ppm ±5 ppm
or 5% rel.
1 ppm 45 s
CO Carbon monoxide, Volume concentration electrochemical sensor 0..10 % 10 ppm
or 0.001 %
±50 ppm
or 5 % rel.
10 ppm 45 s
COmg – Carbon monoxide, mass concentration Calculated from CO volume concentration 0... 1 mg/Nm3 ± 5 mg/Nm3
or 5 % rel.
1 mg/Nm3 45 s
COrel – Carbon monoxide, concentration relative to O2 Calculated from CO and O2 volume concentrations 0... 1 mg/Nm3 ± 5 mg/Nm3
or 5 % rel.
1 mg/Nm3 45 s
NO – Nitric oxide, Volume concentration electrochemical sensor 0..1000 ppm 0.1 or 1ppm
as set
±5 ppm
or 5 % rel.
1 ppm 45 s
NOmg – Nitric oxide, mass concentration Calculated from NO volume concentration 0... 1 mg/Nm3 ±10 mg/Nm3
or 5 % rel.
2 mg/Nm3 45 s
NOrel – Nitric oxide, concentration relative to O2 Calculated from NO and O2 volume concentrations 0... 1 mg/Nm3 ±10 mg/Nm3
or 5 % rel.
2 mg/Nm3 45 s
Optional electrochemical sensors
NO2 – Nitrogen dioxide, Volume concentration electrochemical sensor 0...200 ppm 0.1 or 1 ppm
as set
± 5 ppm
or 5 % rel.
1 ppm 45 s
NO2mg – Nitrogen dioxide, mass concentration Calculated from NO2 volume concentration 0... 1 mg/Nm3 ± 10 mg/Nm3
or 5 % rel.
2 mg/Nm3 45 s
NO2rel - Nitrogen dioxide concentration, relative to O2 Calculated from NO2 and O2 volume concentrations 0... 1 mg/Nm3 ±10 mg/Nm3
or 5 % rel.
2 mg/Nm3 45 s
SO2 – Sulphur dioxide, Volume concentration electrochemical sensor 0...2000 ppm 0.1 or 1 ppm
as set
±5 ppm
or 5 % rel.
1 ppm 45 s
SO2mg – Sulphur dioxide, mass concentration Calculated from SO2 volume concentration 0... 1 mg/Nm3 ± 15 mg/Nm3
or 5 % rel.
3 mg/Nm3 45 s
SO2rel – Sulphur dioxide concentration, relative to O2 Calculated from O2 and SO2 volume concentrations 0... 1 mg/Nm3 ±15 mg/Nm3
or 5 % rel.
3 mg/Nm3 45 s
H2S – Hydrogen sulphide, Volume concentration electrochemical sensor 0...1000 ppm 0.1 or 1 ppm
as set
±5 ppm
or 5 % rel.
1 ppm 45 s
H2Smg – Hydrogen sulphide, mass concentration Calculated from H2S volume concentration 0... 1 mg/Nm3 ±15 mg/Nm3
or 5 % rel.
3 mg/Nm3 45 s
H2Srel – Hydrogen sulphide, mass concentration relative to O2 Calculated from H2S and O2 volume concentrations 0... 1 mg/Nm3 ±15 mg/Nm3
or 5 % rel.
3 mg/Nm3 45 s
H2 –Hydrogen, volume concentration electrochemical sensor 0...2000 ppm 0.1 or 1 ppm
as set
±5 ppm
or 5 % rel.
1 ppm 45 s
H2mg – Hydrogen, mass concentration Calculated from H2 volume concentration 0... 1 mg/Nm3 ±15 mg/Nm3
or 5 % rel.
3 mg/Nm3 45 s
H2rel – Hydrogen, mass concentration relative to O2 Calculated from H2 and O2 volume concentrations 0... 1 mg/Nm3 ±15 mg/Nm3
or 5 % rel.
3 mg/Nm3 45 s
Optional IR sensors
CO – Carbon monoxide, Volume concentration IR Sensor 0..5 % 0...100% 0.01%, 0.1% ±0.3 % abs.,
or 3 % rel.
0.01%, 0.1% 45 s
CO2 – Carbon dioxide, Volume concentration IR Sensor 0..5 % 0...100% 0.01%, 0.1% ±0.3 % abs.,
or 3 % rel.
0.01%, 0.1% 45 s
CH4 – Methane. Volume concentration IR Sensor 0...5%, 0...100% 0.01%, 0.1% ±0.3 % abs.,
or 3 % rel.
0.01%, 0.1% 45 s
Measured temperatures
Tgas – Temperature of the flue gases Thermocouple -20..2000°C 1°C ±2 °C
or 1. 5 % rel.
1 °C 30 s
Tamb – Ambient temperature Thermistor -20..100°C 0.1°C ±1 °C 1 °C 30 s
Further values
Atmospheric pressure DMS bridge 800 hPa ... 1200 hPa 1 hPa ±10 hPa 1 hPa 10 s
Pressure/draught DMS bridge -25 hPa ... +25 hPa 0.1 or 1 Pa as set ±2 Pa
or 5 % rel.
0.1 Pa 10 s
Diff. pressure DMS bridge -20 hPa ... +20 hPa 0.1 or 1 Pa as set ±2 Pa
or 5 % rel.
0.1 Pa 10 s
V – Flow velocity Pitot tube 1..30 m/s 0.1 m/s 0.5 m/s
or 2 % rel.
1 m/s 10 s
Calculated combustion parameters
TI (CO/CO2-Toxic Index) Calculated 0...1 % 0.01 % 0.01 %
Lambda – Excess air Calculated 1..10 0.01 0.01
qA – Combustion loss Calculated 0..100 % 0.1 % 0.1 %
Eta -Efficiency Calculated 0..100 % 0.1 % 0.1 %

Operating data
Parameter Description
Size W x L x H: 600 x 635 x 516 mm
Weight without probe 48 kg
Display size Graphic LCD with backlighting. Contrast set by user, 320 x 240 pixels
Data logger MMC 256 MB
Interface RS232C
Supply Mains 230 VAC 50 ÷ 60Hz
Gas pump Membrane pump, set to 90 l/h
condensate pump peristaltic pump, electronically controlled
Sampling
Probe length 300mm (other lengths available)
Sample hose Heated hose, 3 m
Gas filter Heated filter 20 µm
Analogue outputs
Voltage 8 channels: 0...10 V
Current 8 channels: 0/4...20 mA
Operating temperature 10°C ÷ 40°C
Storage temperature -20 °C ÷ +55 °C
Humidity 5 - 90 %, non-condensing
Humidity 5 - 90 %, non-condensing

IR Sensors
Parameter Description
Measuring method NDIR light absorption, 1 channel with light modulation
Averaging time 2 ÷ 60 s freely programmable
Time T90 < 45 s
Measuring range Begin and end can be freely programmed within the nominal range
Accuracy ±5 ppm abs.,or 5 % rel.
Calibration 10 point calibration, stored in EEPROM
Recalibration not necessary

Electrochemical Sensors

Parameter

Description only for electrochemical sensors
Measuring method electrochemical
Averaging time 2÷30 s freely programmable
Time T90 < 45 s
Measuring range Begin and end can be freely programmed within the nominal range
Accuracy ± 5 ppm or 5 % rel.
Calibration 2 point calibration, variable
Recalibration At least every 12 months