Microprocessor equipped IR Measuring cells
NDIR infrared sensor in flow configuration
IR sensors for measurement of (CO) carbon monoxide, carbon dioxide (CO2) and methane (CH4) in one channel configuration. Available for many measuring ranges from 5 % to 100 % for a variety of industrial applications.
Properties
  • Flow measurement
  • Using microprocessor technology, all on one board
  • High resolution precision measuring technology
  • Short reaction time (T90 < 45 sec.)
  • Permanent automatic sensor check. Need for calibration shown by LED.
  • Temperature compensated
Pictures
Product info
Folder CO2 (54 kB)
Folder CH4 (54 kB)
Control software (25 MB)
Working modes
  • "Smart Zero" - automatic correction of the sensor zero using values from the previous
    1 to 30 days
  • Permanent operation with external zeroing

Standard configuration

  • Linear voltage output - active range (begin and end) may be user-configured within the nominal range 0....10 V.
  • Linear current output - active range (begin and end) may be user-configured within the nominal range 0/4....20 mA
  • Interface RS232C
  • Digital PWM output (0/5 V) - programmable output
  • Zeroing of sensor: manually or externally over isolated contacts
  • Voltage supply 13...30 VDC or 9...24 VAC
Optional equipment
  • Membrane pump SP 550 EC 4
  • USB connection using madur adapter
Data processing and display
  • Measured gas concentration in [ppm] or [%]
    • Measured value can be read digitally over the RS232C
    • Measured value is given as an analog output over the screw connections: 0/4-20 mA and 0-10 V
  • PC program for changing the sensor settings and for data-logging

Operating parameters

Parameter

madIR-D01
Measured gases CO, CO2, CH4
Measuring method infra-red absorption, 1 channel with light modulation
Sampling flow
Operating temperature 0 ÷ +50°C
Storage temperature -20 ÷ +70°C
Humidity 5 ÷ 90%, non-condensing
13...30 VDC or 9...24 VAC
Power max. 3 W
Analog outputs

Begin and end can be user-configured within the range
Current: 0/4 - 20 mA, linear
Voltage: 0 - 10 V, linear

Interface RS 232C
Averaging time 2 ÷ 60 s user-configured
Calibration 10 point calibration stored in EEPROM
Recalibration not necessary, if desired a two point calibration is possible
Abmessungen der Printplatte

105 x 40 x 30 mm

Weight 120 g

Measurement data
Component Range Resolution Minimum value Accuracy Time (T90)
CO - carbon monoxide-volumetric concentration

CO2 - carbon dioxide-volumetric concentration

CH4 - methane volumetric concentration
0...100 % 0.10 % 0.10 % +/-3 % of value,
at least 0.5 % abs.
45 sec.
0....50 % 0.10 % 0.10 % +/-3 % of value,
at least 0.3 % abs.
0...25 % 0.10 % (100 ppm) 0.10 % (100 ppm) +/-3 % of value,
at least 0.15 % abs.
0...10 % 0.10 % (100 ppm) 0.10 % (100 ppm) +/-3 % of value,
at least 0.05 % abs.
0...5 % 0.10 % (100 ppm) 0.10 % (100 ppm) +/-3 % of value,
at least 0.03 % abs.
0...2.5 % 0.001 % (10 ppm) 0.001 % (10 ppm) +/-3 % of value,
at least 0.015 % abs.