madIR-D01 - industrial sensors
NDIR infrared sensor in flow configuration
Microprocessor equipped IR Measuring cells
IR sensors for measurement of carbon dioxide (CO2) and methane (CH4) in one channel configuration. Available for many measuring ranges from 5 % to 100 % for a variety of industrial applications.
Properties
  • Flow or diffusion measurement
  • Using microprocessor technology, all on one board
  • High resolution precision measuring technology
  • Short reaction time (T90 < 45 sec.)
  • Permanent automatic sensor check. Need for calibration shown by LED
  • Temperature compensated
Pictures
Product info
Folder CO2 (54 kB)
Folder CH4 (54 kB)
Control software (25 MB)
Working modes
  • "Smart Zero" - automatic correction of the sensor zero using values from the previous
    1 to 30 days
  • Permanent operation in 3 phases - zero, measure, standby
  • Permanent operation according to timing plan (24 hr. cycle)
  • Permanent operation with external zeroing

Standard configuration

  • ABS housing with all connections
  • Linear voltage output - active range (begin and end) may be user-configured within the nominal range 0....10 V.
  • Linear current output - active range (begin and end) may be user-configured within the nominal range 0/4....20 mA
  • Interface RS232C
  • Digital output (0/5 V) – shows the activity of the sensor (0 V-measuring / 5 V-zeroing or pause)
  • Digital output (0/5 V) – shows the activity of the pump (0 V-still / 5 V-running)
  • Digital PWM output (0/5 V) - programmable output
  • Zeroing of sensor: manually or externally over isolated contacts
  • Voltage supply 13...30 VDC or 9...24 VAC
Optional equipment
  • Display module
  • Membrane pump SP 550 EC 4
  • Data storage on MMC 256MB
  • USB connection using madur adapter
Data processing and display
  • Measured gas concentration in [ppm] or [%]
  • Measured value can be read digitally over the RS232C
  • Measured value can be stored digitally on Memory Card (option)
  • Measured value is given as an analog output over the screw connections: 0/4-20 mA and 0-10 V
  • Measured value can be shown on display (option)
  • PC program for changing the sensor settings and for data-logging

Operating parameters

Parameter

madIR-D01
Measured gases CO, CO2, CH4
Measuring method infra-red absorption, 1 channel with light modulation
Housing ABS
Sampling flow or diffusion
Operating temperature 0 ÷ +50°C
Storage temperature -20 ÷ +70°C
Humidity 5 ÷ 90 %, non-condensing
13...30 VDC or 9...24 VAC
Power approx. 1 W
Analog outputs

Begin and end can be user-configured within the range
Current: 0/4 - 20 mA, linear
Voltage: 0 - 10 V, linear

Interface RS 232C
Averaging time 2 ÷ 60 s user-configured
Calibration 10 point calibration stored in EEPROM
Recalibration not necessary, if desired a two point calibration is possible
Size of PCB

105 x 40 mm

Size of small ABS housing 120 x 80 x 55 mm
Size of large ABS housing

240 x 80 x 55 mm

Weight (small/large housing) 220 g/330 g

Measurement data
Component Range Resolution Minimum value Accuracy Time (T90)
CO - carbon monoxide-volumetric concentration

CO2 - carbon dioxide-volumetric concentration

CH4 - methane volumetric concentration
0...100 % 0.10 % 0.10 % +/-3 % of value,
at least 0.5 % abs.
45 sec.
0....50 % 0.10 % 0.10 % +/-3 % of value,
at least 0.3 % abs.
0...25 % 0.10 % (100 ppm) 0.10 % (100 ppm) +/-3 % of value,
at least 0.15 % abs.
0...10 % 0.10 % (100 ppm) 0.10 % (100 ppm) +/-3 % of value,
at least 0.05 % abs.
0...5 % 0.10 % (100 ppm) 0.10 % (100 ppm) +/-3 % of value,
at least 0.03 % abs.
0...2.5 % 0.001 % (10 ppm) 0.001 % (10 ppm) +/-3 % of value,
at least 0.015 % abs.
A

Copyright © 2007  madur USA. All rights reserved.