| Handheld analyzer using electrochemical sensors. |
|
 |
|
A further development from the tried and tested GA-12 hand-held analyzer. This new analyzer is equipped with the following extra features:
- Can be fitted with up to 4 electrochemical sensors.
- Suitable for a soot test using electronically controlled gas volume.
- Fitted with a powerful Li ion rechargeable battery.
- Big memory capable of storing 32 measurement reports.
An attractive alternative to other, bigger analyzers.
|
|
| Standard configuration |
- Measurement of gas concentrations
- Standard version with two gas sensors: O2, CO
- Two further optional sensors from: NO, NO2, SO2
- CO ambient measurement with a resolution of 0.1 ppm
- Measurement of other parameters
- Measurement of gas and ambient temperatures
- Pressure, draft and differential pressure measurements with 0.1 Pa resolution
- Flow velocity measurement
- Relative humidity measurement probe - optional
- Soot/smoke test according to the Bacharach method - electronically regulated to 1.63 l
- Calculation
- CO2 concentration
- Calcualtion of all relevant combustion parameters
- Processing and presentation of measuring data
- All results shown on display
- Averaging of all measured values. Averaging time: 2, 10, 20, 30, 60, 120, 180 seconds
- Graphical presentation of all measured values as diagram
- Storage for 32 measurements - "measure now, print later"
- Powerful PC program for analyzer settings and data communication
- Software capabilities
- Automatic zeroing when the analyzer is switched on
- All settings for the analyzer carried out with PC programm
- List of 6 common fuels (from DIN)
- 4 further freely programmable fuels
- Permanent automatic check of the instruemnt with acoustic warning and full information in the "control list"
- Hardware capabilities
- Integrated clock/calendar
- IR interface for external printer
- Flexible 24 hours operating time from one charge of batteries
- Full graphic display LCD 66 x 36 mm
- Gas probe with thermocouple and condensate trap
- Power supply for heated probe holder (when powered from the mains)
- RS-232C interface and multifunctional PC program
- Updateable FLASH ROM program memory
- Simple, intuitive operation, like a cell phone
|
|
|
|
|
|
|
|
|
|
|
|
| Optional accessories |
- leather-look case for storage and rubber boot for protection of the instrument
- Magnetic plate to attach the instrument to steel surfaces
- External thermal printer with wireless IR communication
- Optional differential pressure measurement
|
|
|
|
|
|
|
|
|
|
| Parameter |
Measuring method
|
Range |
Resolution |
Accuracy |
Detection
Limit |
Time (T90) |
|
| Gases measured in standard configuration |
|
O2 - oxygen, volumetric concentration
|
electrochemical gas sensor
|
0..25 % |
0.01 % |
±0.2 % or 2 % rel. |
0.01 % |
45 s |
|
CO2 - carbon dioxide, volumetric concentration
|
calculated from volumetric concentration of O2
|
0..25 % |
0.01 % |
±0.2 % or 2 % rel. |
0.01 % |
45 s |
|
CO - ambient carbon monoxide, volumetric concentration
|
electrochemical gas sensor
|
0..100 ppm |
0.1 ppm |
±0.5 ppm
or 5 % rel. |
0.1 ppm |
45 s |
|
CO - carbon monoxide, volumetric concentration
|
electrochemical gas sensor
|
0..20000 ppm |
0.1 or 1 ppm
set by user |
±5 ppm
or 5 % rel. |
1 ppm |
45 s |
|
COmg - carbon monoxide, mass concentration
|
calculated from volumetric concentration of CO
|
0... |
1 mg/Nm3 |
±5 mg/Nm3
or 5 % rel. |
1 mg/Nm3 |
45 s |
|
COrel - carbon monoxide, mass concentration relative to O2
|
calculated from volumetric concentration of CO and O2
|
0... |
1 mg/Nm3 |
±5 mg/Nm3
or 5 % rel. |
1 mg/Nm3 |
45 s |
| Gases measured with optional electrochemical sensors |
|
NO - volumetric concentration of nitrogen oxides
|
electrochemical gas sensor
|
0..1000 ppm |
0.1 or 1 ppm
set by user |
±5 ppm
or 5 % rel. |
1 ppm |
45 s |
|
NOmg - mass concentration of nitrogen oxides
|
calculated from volumetric concentration of NO
|
0... |
1 mg/Nm3 |
±10 mg/Nm3
or 5 % rel. |
2 mg/Nm3 |
45 s |
|
NOrell - mass concentration of nitrogen oxides relative to O2
|
calculated from volumetric concentration of NO and O2
|
0... |
1 mg/Nm3 |
±10 mg/Nm3
or 5 % rel. |
2 mg/Nm3 |
45 s |
|
|
| NO2 - volumetric concentration of nitrogen dioxide. |
electrochemical gas sensor
|
0..1000 ppm
|
1 ppm
|
±5 ppm
or 5 % rel.
|
1 ppm
|
45 s
|
| NO2mg - mass concentration of nitrogen dioxide |
calculated from volumetric concentration of NO2
|
0...
|
1 mg/Nm3
|
±10 mg/Nm3
or 5 % rel.
|
2 mg/Nm3
|
45 s
|
| NO2rel - mass concentration of nitrogen dioxide relative to O2 |
calculated from volumetric concentration of NO2 and O2
|
0...
|
1 mg/Nm3
|
±10 mg/Nm3
or 5 % rel.
|
2 mg/Nm3
|
45 s
|
|
|
| SO2 - volumetric concentration of sulphur dioxide |
electrochemical gas sensor
|
0..5000 ppm
|
1 ppm
|
±5 ppm
or 5 % rel.
|
1 ppm
|
45 s
|
| SO2mg - mass concentration of sulphur dioxide |
calculated from volumetric concentration of SO2
|
0...
|
1 mg/Nm3
|
±15 mg/Nm3
or 5 % rel.
|
3 mg/Nm3
|
45 s
|
| SO2rel - mass concentration of sulphur dioxide relative to O2 |
calculated from volumetric concentration of SO2 and O2
|
0...
|
1 mg/Nm3
|
±15 mg/Nm3
or 5 % rel.
|
3 mg/Nm3
|
45 s
|
| Temperature measurements |
|
Tgas - flue gas temperature
|
Thermocouple
|
-10..1000°C |
0.1 or 1°C
set by user |
±2°C
or 1.5 % rel. |
1°C |
30 s |
|
Tamb - ambient temperature
|
Thermistor
|
-10..100°C |
0.1 or 1°C
set by user |
± 1°C |
1°C |
30 s |
| Calculated parameters |
|
TI (CO/CO2-Toxic Index)
|
calculated
|
0...0.01 |
0.0001 |
0.0001 |
|
|
|
Lambda - excess air number
|
calculated
|
1..10 |
0.01 |
0.01 |
|
|
|
qA - combustion losses
|
calculated
|
0..100 % |
0.1 % |
0.1 % |
|
|
|
Eta - efficiency
|
calculated
|
0..120 % |
0.1 % |
0.1 % |
|
|
| Other measured values |
|
Pressure
|
DMS bridge
|
-25 hPa ... +25 hPa |
0.1 or 1 Pa
set by user |
±2 Pa
or 5 % rel. |
0.1 Pa |
10 s |
|
Diff. Pressure
|
DMS bridge
|
-25 hPa ... +25 hPa |
0.1 or 1 Pa
set by user |
±2 Pa
or 1 % rel. |
0.1 Pa |
10 s |
|
V - flow velocity
|
Pitot tube |
1..30 m/s |
0.1 m/s |
0.5 m/s or 2 % rel. |
1 m/s |
10 s |
|
rH -relative humidity
|
capacitive sensor
|
5..95 % |
0.01 % |
2 % rH or 2 % rel. |
5 % |
10 s |